Lease/Rent

Waters ACQUITY Premier UPLC System with TUV Detector

Listing #:LS-007651
Brand:
Model:186018000, 186018002, 186018004, 186018006
Availability:In Stock
Condition:Refurbished
Warranty:90 Days
Onsite Installation:Available upon request
In Stock

Waters ACQUITY Premier UPLC System with TUV Detector — inert binary UPLC with MaxPeak HPS, Sample Manager FTN, CH-A column heater and TUV detector for high-recovery separations of oligonucleotides, peptides and metal-sensitive compounds. Refurbished and tested by Alliant Scientific.

Description

Waters ACQUITY Premier UPLC with TUV Detector Overview

The Waters ACQUITY Premier UPLC is a binary UPLC system built around MaxPeak High Performance Surfaces, an inert coating on the parts of the flow path that sample and mobile phase touch. Many analytes, including oligonucleotides, phosphorylated compounds, organic acids and chelating drugs, stick to the metal surfaces of a conventional LC, which causes tailing peaks, low recovery and long conditioning runs. The Waters ACQUITY Premier UPLC reduces those losses, so methods for these compounds give sharper peaks and consistent results from the first injection. With a TUV detector, sample manager and column heater, this system is ready for UV methods or to serve as the front end of an LC/MS.

How the ACQUITY Premier Works

The Binary Solvent Manager draws from up to four solvents, two at a time, and mixes the gradient at high pressure, reaching 15,000 psi at flow rates up to 1.0 mL/min. Its gradient delay volume of 90 µL or less means a programmed gradient reaches the column almost immediately, which keeps short UPLC methods short. The Sample Manager FTN injects through a flow-through needle that stays in the flow path, which gives very low carryover, while the sample compartment holds plates or vials at a set temperature. The CH-A column heater pre-heats the mobile phase before the column for stable retention times, and the TUV detector measures absorbance at the selected wavelength as peaks elute.

This Waters ACQUITY Premier UPLC System with TUV Detector is offered refurbished, fully tested, in working condition by Alliant Scientific.

Includes:

  • Waters ACQUITY Premier UPLC System
    • ACQUITY Premier Binary Solvent Manager (P/N 186018000)
    • ACQUITY Premier Sample Manager FTN (P/N 186018002)
    • ACQUITY Premier TUV Detector (P/N 186018006)
    • ACQUITY Premier Column Heater CH-A (P/N 186018004)
    • ACQUITY Premier solvent tray
  • PC loaded with data system
  • 90-day warranty

Browse more refurbished HPLC and UPLC systems

Disclaimer: All brand names and trademarks are the property of their respective owners and are used for identification only. The seller is not authorized by or associated with Waters. Software, if included, is not transferable or upgradable.

Listing last updated: November 2026.

Waters ACQUITY Premier UPLC Specifications

Manufacturer Waters
Model ACQUITY Premier UPLC System with TUV Detector
Condition Refurbished
Flow Path MaxPeak High Performance Surfaces (HPS)
Pump Binary Solvent Manager, high-pressure mixing
Flow Range 0.001–2.000 mL/min
Max Pressure 15,000 psi up to 1.0 mL/min; 9,000 psi up to 2.0 mL/min
Gradient Delay Volume ≤90 µL (FTN configuration)
System Dwell Volume ≤115 µL (FTN configuration)
Solvents Up to 4 (A1/A2, B1/B2)
Sample Manager Flow-Through Needle (FTN)
Injection Volume 0.1–10.0 µL (up to 1000 µL with optional extension loop)
Sample Formats 96/384-well plates, 48-position 2 mL vial plates, microcentrifuge tube plates
Sample Temperature 4.0–40.0 °C
Carryover ≤0.002% (caffeine, UV)
Column Heater CH-A, 20.0–90.0 °C, active pre-heating
Column Capacity One column, up to 4.6 mm I.D. and 150 mm long
Detector ACQUITY Premier TUV
Data System PC loaded with data system included
Warranty 90 days

Additional information

Availability

Brand

Condition

Model

, , ,

Onsite Installation

Warranty

Waters ACQUITY Premier UPLC Key Features

  • MaxPeak HPS flow path: reduces analyte loss and peak tailing from metal surfaces.
  • 15,000 psi binary pump: full sub-2 µm UPLC column performance.
  • ≤90 µL gradient delay volume: fast gradients and short run times.
  • ≤0.002% carryover: flow-through needle injection for clean results between samples.
  • 0.1–10 µL injections: expandable to 1000 µL with an optional loop.
  • 4.0–40.0 °C sample cooling: protects temperature-sensitive samples in the tray.
  • 20.0–90.0 °C column heater: active pre-heating for stable retention times.
  • Plates or vials: 96/384-well plates and 2 mL vial plates.

Waters ACQUITY Premier UPLC Applications

  • Oligonucleotide developers: ion-pair reversed-phase separations of oligonucleotides with better recovery on the inert flow path.
  • Biopharmaceutical labs: UV peptide mapping and protein separations for identity and purity testing.
  • Pharmaceutical QC labs: assay and impurity methods, including chelating and phosphorylated drug compounds.
  • Metabolomics and clinical research labs: organic acids, nucleotides and other metal-sensitive metabolites.
  • Method development groups: fast UPLC gradient screening with low delay volume.
  • LC/MS labs: a low-dispersion, inert UPLC front end for mass spectrometry.

Waters ACQUITY Premier UPLC FAQ

What is MaxPeak HPS?
Waters’ inert High Performance Surfaces in the flow path, which reduce analyte losses to metal surfaces.

What is the maximum pressure?
15,000 psi up to 1.0 mL/min, and 9,000 psi up to 2.0 mL/min.

What is the gradient delay volume?
90 µL or less in the FTN configuration.

What injection volumes can it handle?
0.1 to 10.0 µL, or up to 1000 µL with an optional extension loop.

What is the column temperature range?
20.0 to 90.0 °C.

Can it be used as an LC/MS front end?
Yes. It can feed a mass spectrometer after the TUV detector.

Is a data system included?
Yes. A PC loaded with data system is included.

What warranty is included?
A 90-day warranty.