Waters Acquity UPLC CH-A Column Heater 186015042 — UPLC column heater with active pre-heating, 20 to 90 °C, for ACQUITY UPLC H-Class and I-Class systems. Refurbished and tested by Alliant Scientific.

Description

Waters ACQUITY UPLC CH-A Column Heater Overview

The Waters ACQUITY UPLC CH-A (part number 186015042) is a column heater with active pre-heating for ACQUITY UPLC systems. It holds a single column up to 150 mm long with an internal diameter of 1.0 to 4.6 mm, including a filter or guard column.

Column temperature variations can shift retention times and alter peak shapes. The CH-A controls column temperature from 20 °C, or 5 °C above ambient if higher, up to 90 °C in 0.1 °C increments. Accuracy is ±0.5 °C and stability is ±0.3 °C at the sensor. An active pre-heating device warms the incoming mobile phase to the column’s set temperature before it enters the column. Matching solvent and column temperature minimizes dispersion and supports consistent separation performance from system to system.

The CH-A is designed for use with ACQUITY UPLC H-Class and I-Class systems.

This Waters ACQUITY UPLC CH-A Column Heater is offered fully refurbished and tested, in working condition, by Alliant Scientific.

Includes:

  • Waters ACQUITY UPLC CH-A Column Heater (186015042)
  • 30-Day Warranty

 

Listing last updated: October 2026.

Technical Specifications

Manufacturer Waters
Model ACQUITY UPLC CH-A Column Heater (186015042)
Column Capacity Single column, up to 150 mm length, 1.0 to 4.6 mm I.D., with filter or guard column
Temperature Range 20 °C (or 5 °C above ambient, whichever is higher) to 90 °C
Temperature Setting 0.1 °C increments
Temperature Accuracy ± 0.5 °C at sensor
Temperature Stability ± 0.3 °C at sensor
Solvent Conditioning Active pre-heating
Compatible Systems ACQUITY UPLC H-Class and I-Class
Condition Refurbished and tested
Warranty 30 days

Additional information

Availability

Condition

Brand

Onsite Installation

Model

Warranty

Key Features

  • Active Pre-Heating
    Heats incoming mobile phase to the column set temperature, minimizing dispersion.
  • Wide Temperature Range
    Controls the column from 20 °C (or 5 °C above ambient) to 90 °C.
  • Precise Control
    Settable in 0.1 °C increments with ±0.5 °C accuracy and ±0.3 °C stability at the sensor.
  • Reproducible Retention Times
    Stable column temperature limits shifts in retention time and peak shape.
  • UPLC Column Capacity
    Holds one column up to 150 mm long and 4.6 mm I.D., with filter or guard column.
  • H-Class and I-Class Compatible
    Designed for ACQUITY UPLC H-Class and I-Class systems.

Applications

  • UPLC Method Development
    Use column temperature as a method parameter to tune selectivity and resolution.
  • Routine Quality Control
    Consistent column temperature supports repeatable retention times across batches.
  • Method Transfer
    Matched solvent and column temperature supports consistent performance from system to system.
  • Elevated-Temperature Separations
    Operation up to 90 °C for methods that call for heated columns.

FAQ (Frequently Asked Questions)

What column sizes does the CH-A accept?

A single column up to 150 mm long with an internal diameter of 1.0 to 4.6 mm, including a filter or guard column.

What is the temperature range?

20 °C, or 5 °C above ambient if higher, up to 90 °C, in 0.1 °C increments.

Which systems is the CH-A compatible with?

ACQUITY UPLC H-Class and I-Class systems.

What does active pre-heating do?

It heats the mobile phase to the column’s set temperature before it enters the column, which minimizes dispersion.

Can the CH-A cool a column?

No — it heats only, starting at 5 °C above ambient.

Does this unit include a warranty?

Yes — this unit ships with a 30-day warranty.