LabTech H150-2100LT Low Temperature Recirculating Water Chiller, 230V
Description
For lower temperature requirements, LabTech offers the H150-2100LT recirculating water chiller to ensure accurate and constant cooling conditions for instrumentation.
The chiller is provided with a touchscreen display to monitor and control instrument temperature from -20°C – 35°C and is designed to operate on 230V/50/60Hz with a standard 230V US plug for use with similar electrical requirements. It has a PID Temperature control mode to provide Temperature Stability at +/-0.1℃ with efficient compressor cooling. Cooling capacity for the unit is 2100W at 25℃ and 200W at -15℃ and is ideal for use with rotary evaporation, reflux extraction and AA/ICP applications. The H150-2100LT has a pump capacity of 12.1L/min and a max pressure of 60psi for external circulation. The liquid circuit pressure is monitored and controlled via analog pressure indicator.
Furthermore, the unit is provided with a liquid level indicator for maximum safety of operations. The chiller has a built in drain to easily remove the liquid from the internal reservoir and water connections are located on the rear side.
When setting the H150-2100LT to a lower temperature, a coolant mixture of 50% glycol / 50% water should be used to ensure proper functioning of the chiller.
This is a Brand‑new LabTech H150-2100LT Low Temperature Recirculating Water Chiller, 230V , ready for laboratory use.
Alliant Scientific is an authorized LabTech distributor.
H150-2100LT (230V)Â
- Part No.              LW2100LT
- Temp. Range          -20℃~35℃
- Temp. Stability        ±0.1℃
- Temp. Control Mode   PID
- Cooling Capacity       2100W@25℃
-                       200W@-15℃
- Pump Flow            12.1L/min
- L×W×H                29×18×26 inches
- Weight                203 lbs
- Power Supply          230V,50/60Hz
Additional information
| Brand | |
|---|---|
| Condition | |
| Model | |
| Warranty | |
| Availability | |
| Onsite Installation |